16-channel bow tie antenna transceiver array for cardiac MR at 7.0 Tesla

Autor/innen

  • C. Oezerdem
  • L. Winter
  • A. Graessl
  • K. Paul
  • A. Els
  • O. Weinberger
  • J. Rieger
  • A. Kuehne
  • M. Dieringer
  • F. Hezel
  • D. Voit
  • J. Frahm
  • T. Niendorf

Journal

  • Magnetic Resonance in Medicine

Quellenangabe

  • Magn Reson Med 75 (6): 2553-2565

Zusammenfassung

  • PURPOSE: To design, evaluate, and apply a bow tie antenna transceiver radiofrequency (RF) coil array tailored for cardiac MRI at 7.0 Tesla (T). METHODS: The radiofrequency (RF) coil array comprises 16 building blocks each containing a bow tie shaped lambda/2-dipole antenna. Numerical simulations were used for transmission field homogenization and RF safety validation. RF characteristics were examined in a phantom study. The array's suitability for high spatial resolution two-dimensional (2D) CINE imaging and for real time imaging of the heart was examined in a volunteer study. RESULTS: The arrays transmission fields and RF characteristics are suitable for cardiac MRI at 7.0T. The coil performance afforded a spatial resolution as good as (0.8 x 0.8 x 2.5) mm3 for segmented 2D CINE MRI at 7.0T which is by a factor of 12 superior versus standardized protocols used in clinical practice at 1.5T. The proposed transceiver array supports 1D acceleration factors of up to R = 6 without impairing image quality significantly. CONCLUSION: The 16-channel bow tie antenna transceiver array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0 Tesla.


DOI

doi:10.1002/mrm.25840